Evaluation of Chiron HIV-1/HIV-2 recombinant immunoblot assay JOURNAL OF CLINICAL MICROBIOLOGY Kline, R. L., McNairn, D., Holodniy, M., Mole, L., Margolis, D., Blattner, W., Quinn, T. C. 1996; 34 (11): 2650-2653

Abstract

In a study to determine the reliability, sensitivity, and specificity of the Chiron RIBA HIV-1/HIV-2 Strip Immunoblot Assay (RIBA HIV-1/2 SIA) for confirmation of human immunodeficiency virus type 1 (HIV-1) and HIV-2 antibodies, 1,263 serum samples from various populations in the United States, Caribbean, Africa, India, and Thailand were evaluated by RIBA HIV-1/2 SIA, and the results were compared with those obtained by an HIV-1 Western blot (immunoblot) assay. All sera were tested by HIV enzyme immunoassay, RIBA HIV-1/2 SIA, and Western blotting. Samples with discrepant results were further tested by an HIV-1 and/or HIV-2 immunofluorescent-antibody assay and HIV-1 p24 antigen assay. The RIBA HIV-1/2 SIA detected all 17 HIV-1 and HIV-2 dually reactive serum samples, all 215 HIV-2-positive serum samples, and 480 of 481 HIV-1-positive serum samples for a sensitivity of 99.8%. Of 548 negative samples, 523 were RIBA HIV-1/2 SIA negative, for a specificity of 95.4%, with 22 (4%) samples interpreted as indeterminate and 3 (0.6%) interpreted as falsely positive. Western blotting detected 391 of 548 negative samples (specificity, 71.4%), with 152 (27.7%) samples interpreted as indeterminate and 5 (0.9%) interpreted as falsely positive. In conclusion, the RIBA HIV-1/2 SIA had a sensitivity comparable to that of Western blotting and could discriminate HIV-1 from HIV-2 in one blot, providing a cost advantage. Because of its high degree of specificity, the RIBA HIV-1/2 SIA further reduced the number of indeterminate results found by Western blotting, providing a more accurate means of assessing seronegative individuals.

View details for Web of Science ID A1996VM55600003

View details for PubMedID 8897158