Power Doppler imaging in preoperative planning and postoperative monitoring of muscle flaps JOURNAL OF CLINICAL ULTRASOUND Loh, N. N., Chen, I. Y., Olcott, E., Jeffrey, R. B., Hui, K. C., PERSONS, B., Lineaweaver, W. C. 1997; 25 (9): 465-471

Abstract

We assessed the utility of power Doppler imaging (PDI) in preoperative planning and postoperative evaluation of microvascular tissue transfers.Twenty-five PDI studies were performed on 23 patients using a 5-10-MHz linear-array transducer. Thirteen patients were assessed preoperatively for patency of the desired donor vessel; 8 of them had surgical scars overlying the desired vascular territory. Twelve patients (including 2 from the first group) were evaluated postoperatively for patency of the vascular anastomoses and adequacy of the blood supply to the transferred tissue.Twelve of the 13 patients assessed preoperatively had successful flap transfers. Four of the 8 patients with scars over the desired vascular territories had absent or aberrant arteries, necessitating a change in the operative plan. None of these patients had operative complications. Eight of the 12 patients scanned postoperatively had patent anastomoses. In 2 of these patients, impending surgery was averted when the adequacy of the tissue blood supply was established with PDI. In 4 patients, PDI showed arterial or venous compromise, which was confirmed at surgery.PDI is a useful technique in microsurgical tissue transfer for assessing the patency of desired donor vessels preoperatively and for postoperative evaluation of blood supply.

View details for Web of Science ID A1997YC38100001

View details for PubMedID 9350564