In vivo real-time intravascular MRI JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE Rivas, P. A., Nayak, K. S., Scott, G. C., McConnell, M. V., Kerr, A. B., Nishimura, D. G., Pauly, J. M., Hu, B. S. 2002; 4 (2): 223-232

Abstract

The Magnetic resonance imaging (MRI) is an emerging technology for catheter-based imaging and interventions. Real-time MRI is a promising methodfor overcoming catheter and physiologic motion for intravascular imaging.All imaging was performed on a 1.5 T Signa MRI scanner with high-speed gradients. Multiple catheter coils were designed and constructed, including low-profile, stub-matched coils. Coil sensitivity patterns and SNR measurements were compared. Real-time imaging was performed with an interleaved spiral sequence using a dedicated workstation, providing real-time data acquisition, image reconstruction and interactive control and display. Real-time "black-blood" imaging was achieved through incorporation of off-slice saturation pulses. The imaging sequence was tested in a continuous flow phantom and then in vivo in the rabbit aorta using a 2 mm catheter coil.The real-time intravascular imaging sequence achieved 120-440 micron resolution at up to 16 frames per second. Low-profile stub-tuned catheter coils achieved similar SNR to larger traditional coil designs. In the phantom experiments, addition of real-time black-blood saturation pulses effectively suppressed the flow signal and allowed visualization of the phantom wall. In vivo experiments clearly showed real-time intravascular imaging of the rabbit aortic wall with minimal motion artifacts and effective blood signal suppression.Real-time imaging with low-profile coil designs provides significant enhancements to intravascular MRI.

View details for PubMedID 12074137