The national average for achieving the KDOQI-recommended hemoglobin (Hgb) target level of 11-12 g/dL is low with the current anemia management protocol of measuring Hgb levels every 2-4 weeks to guide intervention. The objective of this study was to correlate initial Hgb readings from the CRIT-LINE monitor with actual serum Hgb levels in pediatric patients on hemodialysis (HD).Data were collected from pediatric HD patients who had Hgb tests ordered for routine and/or clinical reasons. Hgb concentrations were read with the CRIT-LINE after 0.5 or 1 L of blood had been processed by HD in patients with a body weight of =20 or >20 kg, respectively. Ultrafiltration was kept at a minimum until the CRIT-LINE Hgb was read.In total, 217 Hgb readings from 23 HD patients were analyzed. Results showed a statistically significant correlation between CRIT-LINE readings and laboratory Hgb measurements (r?=?0.94, p?