Spiral magnetic resonance coronary angiography - Direct comparison of 1.5 tesla vs. 3 tesla JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE Yang, P. C., Nguyen, P., Shimakawa, A., Brittain, J., Pauly, J., Nishimura, D., Hu, B., McConnell, M. 2004; 6 (4): 877-884

Abstract

MR coronary angiography (MRCA) has been demonstrated successfully at 3 Tesla (T). However, the advantages remain unclear. No systematic comparison of MRCA between 1.5 T and 3 T has been performed. Therefore, anatomic coverage, image quality, signal-to-noise ratio (SNR), contrast-to-noise ration (CNR), and susceptibility artifacts were compared in 23 subjects.Identical real-time (RT) and high-resolution (HR) sequences were implemented on the GE 1.5 T (Signa Twinspeed) and 3.0 T (Signa VH/i) whole body systems (GE, Milwaukee, WI). Both scanners were equipped with high-performance gradient systems capable of 40 mT/m peak amplitude and 150 mT/m/ms slew rate. Real-time localization of the coronary arteries was followed by a cardiac-gated, breath-hold HR sequence. Twenty-three subjects were recruited consecutively and underwent both 3 T and 1.5 T MRCA within one week. Coronary coverage based on the number of coronary segments visualized, image quality using a grading scale, SNR, CNR, and presence of susceptibility artifacts were analyzed. A significant improvement in SNR (47%), CNR (30%), and image quality were seen in 3 T. However, a significant increase in susceptibility artifacts was also noted.MRCA at 3 T significantly improves SNR, CNR, and image quality at the expense of susceptibility artifacts. Further optimization of the imaging parameters at 3 T may facilitate clinical implementation of MRCA.

View details for DOI 10.1081/JCMR.20036180

View details for PubMedID 15646891