Infection and Erosion Rates in Trials of a Cranially Implanted Neurostimulator Do Not Increase with Subsequent Neurostimulator Placements STEREOTACTIC AND FUNCTIONAL NEUROSURGERY Weber, P. B., Kapur, R., Gwinn, R. P., Zimmerman, R. S., Courtney, T. A., Morrell, M. J. 2017; 95 (5): 325–29

Abstract

The RNS® System utilizes a cranially implanted neurostimulator attached to leads placed at the seizure focus to provide brain responsive stimulation for the treatment of medically intractable partial onset epilepsy. Infection and erosion rates related to the cranial implant site were assessed overall and by neurostimulator procedure to determine whether rates increased with additional procedures.Infection and erosion rates were calculated as (1) chance per neurostimulator procedure, (2) incidence per patient implant year, and (3) rates for initial and each subsequent neurostimulator implant (generalized estimating equation).In 256 patients followed for an average of 7 years, the infection rate was 3.7% per neurostimulator procedure (n = 31/840), and the rate of erosions was 0.8% per neurostimulator procedure (n = 7/840). Rates did not increase with subsequent neurostimulator procedures (p = 0.66, infection; p = 0.70, erosion). A prior infection or erosion at the implant site did not significantly increase the risk at a later procedure (p = 0.05 for all combinations).These data indicate that the risk for infection compares favorably to other neurostimulation devices and suggest that rates of infection and erosion do not increase with subsequent neurostimulator replacements.

View details for DOI 10.1159/000479288

View details for Web of Science ID 000415150700005

View details for PubMedID 28910805

View details for PubMedCentralID PMC5804848