Rigid Motion Correction for Brain PET/MR Imaging using Optical Tracking. IEEE transactions on radiation and plasma medical sciences Spangler-Bickell, M. G., Khalighi, M. M., Hoo, C., DiGiacomo, P. S., Maclaren, J., Aksoy, M., Rettmann, D., Bammer, R., Zaharchuk, G., Zeineh, M., Jansen, F. 2019; 3 (4): 498-503

Abstract

A significant challenge during high-resolution PET brain imaging on PET/MR scanners is patient head motion. This challenge is particularly significant for clinical patient populations who struggle to remain motionless in the scanner for long periods of time. Head motion also affects the MR scan data. An optical motion tracking technique, which has already been demonstrated to perform MR motion correction during acquisition, is used with a list-mode PET reconstruction algorithm to correct the motion for each recorded event and produce a corrected reconstruction. The technique is demonstrated on real Alzheimer's disease patient data for the GE SIGNA PET/MR scanner.

View details for DOI 10.1109/TRPMS.2018.2878978

View details for PubMedID 31396580

View details for PubMedCentralID PMC6686883