Novel Neonatal Umbilical Catheter Protection and Stabilization Device in In vitro Model of Catheterized Human Umbilical Cords: Effect of Material and Venting on Bacterial Colonization. American journal of perinatology Wood, L. S., Fuerch, J. H., Dambkowski, C. L., Chehab, E. F., Torres, S., Shih, J. D., Venook, R., Wall, J. K. 2019

Abstract

OBJECTIVE: Umbilical central lines deliver life-saving medications and nutrition for neonates; however, complications associated with umbilical catheters (UCs) occur more frequently than in adults with central lines (i.e., line migration, systemic infection). We have developed a device for neonatal UC protection and stabilization to reduce catheter exposure to bacteria compared with the standard of care: "goal post" tape configuration. This study analyzes the effect of device venting and material on bacterial load of human umbilical cords in vitro.STUDY DESIGN: Catheters were inserted into human umbilical cord segments in vitro, secured with plastic or silicone vented prototype versus tape, and levels of bacterial colonization were compared between groups after 7 days of incubation.RESULTS: Nonvented plastic prototype showed increased bacterial load compared with goal post (p=0.04). Colonization was comparable between the goal post and all vented plastic prototypes (p=0.30) and when compared with the vented silicone device (p=1).CONCLUSION: A novel silicone device does not increase external bacterial colonization compared with the current standard of care for line securement, and may provide a safe, convenient alternative to standard adhesive tape for UC stabilization. Future studies are anticipated to establish safety in vivo, alongside benefits such as migration and infection reduction.

View details for DOI 10.1055/s-0039-1700542

View details for PubMedID 31739365