Real-Time Semi-Automated and Automated Voxel Placement using fMRI Targets for Repeated Acquisition Magnetic Resonance Spectroscopy.
Real-Time Semi-Automated and Automated Voxel Placement using fMRI Targets for Repeated Acquisition Magnetic Resonance Spectroscopy. Journal of neuroscience methods Bishop, J. H., Geoly, A., Khan, N., Tischler, C., Krueger, R., Keshava, P., Amin, H., Baltusis, L., Wu, H., Spiegel, D., Williams, N., Sacchet, M. D. 2023: 109853Abstract
BACKGROUND: Currently, magnetic resonance spectroscopy (MRS) is dependent on the investigative team to manually prescribe, or demarcate, the desired tissue volume-of-interest. The need for a new method to automate precise voxel placements is warranted to improve the utility and interpretability of MRS data.NEW METHOD: We propose and validate robust and real-time methods to automate MRS voxel placement using functionally defined coordinates within the prefrontal cortex. Data were collected and analyzed using two independent prospective studies: 1) two independent imaging days with each consisting of a multi-session sandwich design (MRS data only collected on one of the days determined based on scan time) and 2) a longitudinal design. Participants with fibromyalgia syndrome (N=50) and major depressive disorder (N=35) underwent neuroimaging. MRS acquisitions were acquired at 3-tesla. Evaluation of the reproducibility of spatial location and tissue segmentation was assessed for: 1) manual, 2) semi-automated, and 3) automated voxel prescription approaches RESULTS: Variability of voxel grey and white matter tissue composition was reduced using automated placement protocols. Spatially, post- to pre-voxel center-of-gravity distance was reduced and voxel overlap increased significantly across datasets using automated compared to manual procedures COMPARISON WITH EXISTING METHODS: Manual prescription, the current standard in the field, can produce inconsistent data across repeated acquisitions. Using automated voxel placement, we found reduced variability and more consistent voxel placement across multiple acquisitions CONCLUSIONS: These results demonstrate the within subject reliability and reproducibility of a method for reducing variability introduced by spatial inconsistencies during MRS acquisitions. The proposed method is a meaningful advance toward improved consistency of MRS data in neuroscience and can be utilized for multi-session and longitudinal studies.
View details for DOI 10.1016/j.jneumeth.2023.109853
View details for PubMedID 37031764