Prevention of Nocturnal Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension DIABETES CARE Buckingham, B., Chase, H. P., Dassau, E., Cobry, E., Clinton, P., Gage, V., Caswell, K., Wilkinson, J., Cameron, F., Lee, H., Bequette, B. W., Doyle, F. J. 2010; 33 (5): 1013-1017

Abstract

The aim of this study was to develop a partial closed-loop system to safely prevent nocturnal hypoglycemia by suspending insulin delivery when hypoglycemia is predicted in type 1 diabetes.Forty subjects with type 1 diabetes (age range 12-39 years) were studied overnight in the hospital. For the first 14 subjects, hypoglycemia (<60 mg/dl) was induced by gradually increasing the basal insulin infusion rate (without the use of pump shutoff algorithms). During the subsequent 26 patient studies, pump shutoff occurred when either three of five (n = 10) or two of five (n = 16) algorithms predicted hypoglycemia based on the glucose levels measured with the FreeStyle Navigator (Abbott Diabetes Care).The standardized protocol induced hypoglycemia on 13 (93%) of the 14 nights. With use of a voting scheme that required three algorithms to trigger insulin pump suspension, nocturnal hypoglycemia was prevented during 6 (60%) of 10 nights. When the voting scheme was changed to require only two algorithms to predict hypoglycemia to trigger pump suspension, hypoglycemia was prevented during 12 (75%) of 16 nights. In the latter study, there were 25 predictions of hypoglycemia because some subjects had multiple hypoglycemic events during a night, and hypoglycemia was prevented for 84% of these events.Using algorithms to shut off the insulin pump when hypoglycemia is predicted, it is possible to prevent hypoglycemia on 75% of nights (84% of events) when it would otherwise be predicted to occur.

View details for DOI 10.2337/dc09-2303

View details for PubMedID 20200307