Imaging of the Wrist at 1.5 Tesla Using Isotropic Three-Dimensional Fast Spin Echo Cube JOURNAL OF MAGNETIC RESONANCE IMAGING Stevens, K. J., Wallace, C. G., Chen, W., Rosenberg, J. K., Gold, G. E. 2011; 33 (4): 908-915

Abstract

To compare three-dimensional fast spin echo Cube (3D-FSE-Cube) with conventional 2D-FSE in MR imaging of the wrist.The wrists of 10 volunteers were imaged in a 1.5 Tesla MRI scanner using an eight-channel wrist coil. The 3D-FSE-Cube images were acquired in the coronal plane with 0.5-mm isotropic resolution. The 2D-FSE images were acquired in both coronal and axial planes for comparison. An ROI was placed in fluid, cartilage, and muscle for SNR analysis. Comparable coronal and axial images were selected for each sequence, and paired images were randomized and graded for blurring, artifact, anatomic details, and overall image quality by three blinded musculoskeletal radiologists.SNR of fluid, cartilage and muscle at prescribed locations were higher using 3D-FSE-Cube, without reaching statistical significance. Fluid-cartilage CNR was also higher with 3D-FSE-Cube, but not statistically significant. Blurring, artifact, anatomic details, and overall image quality were significantly better on coronal 3D-FSE-Cube images (P < 0.001), but significantly better on axial 2D-FSE images compared with axial 3D-FSE-Cube reformats (P < 0.01).Isotropic data from 3D-FSE-Cube allows reformations in arbitrary scan planes, which may make multiple 2D acquisitions unnecessary, and improve depiction of complex wrist anatomy. However, axial reformations suffer from blurring, likely due to T2 decay during the long echo train, limiting overall image quality in this plane.

View details for DOI 10.1002/jmri.22494

View details for PubMedID 21448957