Subject-specific models of susceptibility-induced B0 field variations in breast MRI JOURNAL OF MAGNETIC RESONANCE IMAGING Jordan, C. D., Daniel, B. L., Koch, K. M., Yu, H., Conolly, S., Hargreaves, B. A. 2013; 37 (1): 227-232

Abstract

To rapidly calculate and validate subject-specific field maps based on the three-dimensional shape of the bilateral breast volume.Ten healthy female volunteers were scanned at 3 Tesla using a multi-echo sequence that provides water, fat, in-phase, out-of-phase, and field map images. A shape-specific binary mask was automatically generated to calculate a computed field map using a dipole field model. The measured and computed field maps were compared by visualizing the spatial distribution of the difference field map, the mean absolute error, and the 80% distribution widths of frequency histograms.The 10 computed field maps had a mean absolute error of 38 Hz (0.29 ppm) compared with the measured field maps. The average 80% distribution widths for the histograms of all of the computed, measured, and difference field maps are 205 Hz, 233 Hz, and 120 Hz, respectively.The computed field maps had substantial overall agreement with the measured field maps, indicating that breast MRI field maps can be computed based on the air-tissue interfaces. These estimates may provide a predictive model for field variations and thus have the potential to improve applications in breast MRI.

View details for DOI 10.1002/jmri.23762

View details for Web of Science ID 000312720000025

View details for PubMedID 22865658