THERMODILUTION CARDIAC-OUTPUT MEASUREMENT - EFFECTS OF THE RESPIRATORY CYCLE ON ITS REPRODUCIBILITY JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION Stevens, J. H., Raffin, T. A., Mihm, F. G., Rosenthal, M. H., STETZ, C. W. 1985; 253 (15): 2240-2242

Abstract

Thermodilution cardiac output measurements are commonly employed in the management of critically ill patients. Serial measurements often show significant variation, and poor reproducibility limits their clinical utility. There are no clinical studies revealing when to perform thermodilution cardiac output measurements in relation to the respiratory cycle. We prospectively studied 32 patients in a randomized scheme comparing three thermodilution cardiac output measurements at peak-inspiration, at end-exhalation, or randomly in spontaneously breathing and mechanically ventilated patients. Saline injections initiated at peak-inspiration or end-exhalation resulted in cardiac output measurements with much smaller standard deviations than those seen with random injections. Thermodilution cardiac output measurements performed at random times in the respiratory cycle should be avoided, and we recommend initiating these measurements at end-exhalation.

View details for Web of Science ID A1985AFE4400026

View details for PubMedID 3974116