Bispectral index (BIS) and burst suppression: revealing a part of the BIS algorithm. Journal of clinical monitoring and computing Bruhn, J., Bouillon, T. W., Shafer, S. L. 2000; 16 (8): 593-596

Abstract

The bispectral index (BIS) is a complex EEG parameter which integrates several disparate descriptors of the EEG into a single variable. One of the subparameters incorporated in the BIS is the suppression ratio, quantifying the percentage of suppression during burst suppression pattern. The exact algorithm used to synthetize the information to the BIS value is unpublished and still unknown. This study provides insight into the integration of the suppression ratio into the BIS algorithm.EEG data of 10 healthy volunteers during propofol infusion were analyzed. Propofol concentrations were ramped up to 4 predetermined concentrations (1, 2, 3, 4, 6, 8, 9, or 12 microg/ml) using a computer controlled infusion pump (STANPUMP). EEG recordings were performed with an Aspect A-1000 EEG monitor (Version 3.22). The relationship of the processed EEG variables bispectral index and suppression ratio, calculated by the Aspect A-1000 monitor, was analyzed.Up to 40% suppression ratio the average BIS values remained constant regardless of suppression ratios (r = 0.13). Beyond a suppression ratio of 40%, BIS and suppression ratio were invariably linearly correlated (r = -1). At a suppression ratio > or = 40% the BIS value could be calculated as BIS = 50 - suppression ratio/2.Suppression ratio values > 40% are linearly correlated with BIS values from 30 to 0. An increasing anesthetic drug effect resulting in an increase of the duration of suppression to a suppression ratio up to 40% is not adequately reflected by the BIS value.

View details for PubMedID 12580235