Using a dedicated lung-marker system for localization of pulmonary nodules before thoracoscopic surgery AMERICAN JOURNAL OF ROENTGENOLOGY Partik, B. L., Leung, A. N., Muller, M. R., Breitenseher, M., Eckersberger, F., Dekan, G., Helbich, T. H., Metz, V. 2003; 180 (3): 805-809

Abstract

Our aim was to evaluate the effectiveness of a commercially available dedicated lung-marker system for localization of pulmonary nodules before video-assisted thoracoscopic surgery.Guidewires were positioned under CT fluoroscopy guidance in 16 patients (11 men, five women; age range, 39-79 years; mean age, 60.4 years). We measured the size of the targeted nodule, its distance to the closest pleural surface, the angle between the introducer needle and the chest wall, and the time for performance of the procedure in each patient. Note was made of any complications after guidewire placement.In the 16 patients, the average nodule size was 6.7 mm (range, 3-12 mm), the average distance to the pleural surface was 10.6 mm (range, 3-22 mm), and the average pleural puncture angle was 59 degrees (range, 25-78 degrees). The marking procedure was completed within an average of 9.5 min (range, 7-15 min). Small pneumothoraces occurred in five (31.3%) of 16 patients. In 15 (93.8%) of 16 patients, thoracoscopic resection of the targeted nodule was successful; in one patient with dyspnea (6.3%), inaccurate localization resulting in an open thoracotomy occurred because an intervening fissure was not visualized. Dislodgement of the guidewire into the pleural space occurred in one patient (6.3%).The dedicated lung-marker system is a fast and effective method for localization of pulmonary nodules before thoracoscopic resection.

View details for Web of Science ID 000181153400042

View details for PubMedID 12591700