Case report: Pseudotumor associated with corrosion of a femoral component with a modular neck and a ceramic-on-polyethylene bearing. Journal of long-term effects of medical implants Messana, J., Adelani, M., Goodman, S. B. 2014; 24 (1): 1-5

Abstract

Pseudotumor is a rare complication that can occur following hip arthroplasty. This complication may present with pain, swelling, and decreased function and may lead to bone and soft-tissue destruction. We report a case of pseudotumor formation resulting from corrosion of a modular neck in a hip replacement with a ceramic-on-polyethylene bearing. The patient underwent successful revision surgery using an extended trochanteric osteotomy, removal of the entire stem, and implantation of a new femoral stem and ceramic-polyethylene bearing without a modular neck.

View details for PubMedID 24941400