Oxford meniscal bearing knee versus the Marmor knee in unicompartmental arthroplasty for arthrosis - A Swedish multicenter survival study JOURNAL OF ARTHROPLASTY Lewold, S., Goodman, S., Knutson, K., Robertsson, O., Lidgren, L. 1995; 10 (6): 722-731

Abstract

In the Swedish Knee Arthroplasty Study, all 699 Oxford meniscal bearing cemented unicompartmental prostheses (Biomet, Bridgend, UK) were identified and analyzed regarding failure pattern and compared with all Marmor prostheses (Smith & Nephew Richards, Orthez, France) and with a time-, age-, and sex-matched subset of Marmor prostheses using survival statistics expressed as cumulative revision rates. After 1 year there was already a higher rate, and after 6 years the rate of the Oxford group was more than twice that of the Marmor group. There were 50 revisions in the Oxford group: dislocating meniscus in 16, loosening of the femoral component in 6, tibial component in 4, both components in 4, contralateral arthrosis in 10, infection in 4, and technical failure with instability, pain, and/or impingement of the meniscal bearing anterior in the femoral condyle in 6. It is still unclear if the design with the sliding menisci will, in the long turn, reduce wear and loosening, thereby compensating for the initially inferior results. It is recommended that until this question is clarified, the Oxford knee should be used on a limited scale for long-term comparative studies only.

View details for Web of Science ID A1995TL44800003