Learn about the flu shot, COVID-19 vaccine, and our masking policy »
New to MyHealth?
Manage Your Care From Anywhere.
Access your health information from any device with MyHealth. You can message your clinic, view lab results, schedule an appointment, and pay your bill.
ALREADY HAVE AN ACCESS CODE?
DON'T HAVE AN ACCESS CODE?
NEED MORE DETAILS?
MyHealth for Mobile
Get the iPhone MyHealth app »
Get the Android MyHealth app »
Abstract
Exercise-associated hypoglycemia is a common adverse event in people with type 1 diabetes. Previous in silico testing by our group demonstrated superior exercise-associated hypoglycemia mitigation when a predictive low glucose suspend (PLGS) algorithm was augmented to incorporate activity data. The current study investigates the effectiveness of an accelerometer-augmented PLGS algorithm in an outpatient exercise protocol. Subjects with type 1 diabetes on insulin pump therapy participated in two structured soccer sessions, one utilizing the algorithm and the other using the subject's regular basal insulin rate. Each subject wore their own insulin pump and a Dexcom G4™ Platinum continuous glucose monitor (CGM); subjects on-algorithm also wore a Zephyr BioHarness™ 3 accelerometer. The algorithm utilized a Kalman filter with a 30-minute prediction horizon. Activity and CGM readings were manually entered into a spreadsheet and at five-minute intervals, the algorithm indicated whether the basal insulin infusion should be on or suspended; any changes were then implemented by study staff. The rate of hypoglycemia during and after exercise (until the following morning) was compared between groups. Eighteen subjects (mean age 13.4 ± 3.7 years) participated in two separate sessions 7-22 days apart. The difference in meter blood glucose levels between groups at each rest period did not achieve statistical significance at any time point. Hypoglycemia during the session was recorded in three on-algorithm subjects, compared to six off-algorithm subjects. In the postexercise monitoring period, hypoglycemia occurred in two subjects who were on-algorithm during the session and four subjects who were off-algorithm. The accelerometer-augmented algorithm failed to prevent exercise-associated hypoglycemia compared to subjects on their usual basal rates. A larger sample size may have achieved statistical significance. Further research involving an automated system, a larger sample size, and an algorithm design that favors longer periods of pump suspension is necessary.
View details for DOI 10.1177/1932296814551045
View details for PubMedID 25231116