Tracheal width and teft double-lumen tube size: a formula to estimate teft-bronchial width JOURNAL OF CLINICAL ANESTHESIA Brodsky, J. B., Lemmens, H. J. 2005; 17 (4): 267-270

Abstract

To determine which patient parameters best predict left bronchial width (LBW) when selecting the correct size double-lumen tube (DLT). If LBW is known, a DLT that will fit that bronchus can be chosen.Prospective study.University medical center.Three hundred twenty-one consecutive patients scheduled for thoracic surgery and for whom there was a chest radiograph and for whom tracheal width (TW) and LBW could be measured.Tracheal width and LBW were directly measured from the chest radiograph. Patient demographic data were recorded and then analyzed to see which factor(s) best predicted LBW. Parameters often used for DLT selection (age, sex, height, and weight) as well as TW were compared by univariate and multivariate statistical analysis to see which factor(s) most accurately predicted LBW.There were weak but significant correlations between age and height and LBW in men, and height and LBW in women. Multivariate statistical analysis showed that, for both men and women, TW was the best predictor of LBW. Sex, height, and weight did not improve predictability over TW alone. The equation that best predicts LBW for both sexes is: LBWmm = (0.50)(TWmm) + 3.7 mm. This model explains 46% of the variance in LBW. As structures measured from a chest radiograph are magnified by 10%, the formula to predict LBW, which normalizes for this magnification factor, is: LBWmm = (0.45)(TWmm(CXR)) + 3.3 mm.Direct airway measurement is the most accurate way to select an appropriate DLT. However, when direct measurement of LBW cannot be performed, estimating LBW from TW is a better predictor of LBW than either sex, height, or weight.

View details for DOI 10.1016/j.jclinane.2004.07.008

View details for Web of Science ID 000230236200006

View details for PubMedID 15950850