3D Cartesian MRI with compressed sensing and variable view sharing using complementary poisson-disc sampling
3D Cartesian MRI with compressed sensing and variable view sharing using complementary poisson-disc sampling MAGNETIC RESONANCE IN MEDICINE Levine, E., Daniel, B., Vasanawala, S., Hargreaves, B., Saranathan, M. 2017; 77 (5): 1774-1785Abstract
To enable robust, high spatio-temporal-resolution three-dimensional Cartesian MRI using a scheme incorporating a novel variable density random k-space sampling trajectory allowing flexible and retrospective selection of the temporal footprint with compressed sensing (CS).A complementary Poisson-disc k-space sampling trajectory was designed to allow view sharing and varying combinations of reduced view sharing with CS from the same prospective acquisition. These schemes were used for two-point Dixon-based dynamic contrast-enhanced MRI (DCE-MRI) of the breast and abdomen. Results were validated in vivo with a novel approach using variable-flip-angle data, which was retrospectively accelerated using the same methods but offered a ground truth.In breast DCE-MRI, the temporal footprint could be reduced 2.3-fold retrospectively without introducing noticeable artifacts, improving depiction of rapidly enhancing lesions. Further, experiments with variable-flip-angle data showed that reducing view sharing improved accuracy in reconstruction and T1 mapping. In abdominal MRI, 2.3-fold and 3.6-fold reductions in temporal footprint allowed reduced motion artifacts.The complementary-Poisson-disc k-space sampling trajectory allowed a retrospective spatiotemporal resolution tradeoff using CS and view sharing, imparting robustness to motion and contrast enhancement. The technique was also validated using a novel approach of fully acquired variable-flip-angle acquisition. Magn Reson Med, 2016. © 2016 Wiley Periodicals, Inc.
View details for DOI 10.1002/mrm.26254
View details for Web of Science ID 000399666400005
View details for PubMedCentralID PMC5074926