Cerebral Oximetry Fails as a Monitor of Brain Perfusion in Cardiac Surgery: A Case Report. A&A practice McAvoy, J., Jaffe, R., Brock-Utne, J., Lopez, J., Brodt, J. 2019

Abstract

Cerebral oximetry is commonly being advocated as a monitor for regional cerebral tissue oxygenation during cardiac surgery. We have increasing concern about the accuracy of this monitor, including the current systems entering the market, with new probes and algorithms. We present 2 cases where cerebral oximetry failed to accurately portray cerebral oxygenation. In the current form, cerebral oximetry may at best be an expensive tool without any benefit on outcomes. In addition, it may contribute to misleading and confusing clinical data.

View details for PubMedID 30663993