A preliminary report of gonadal-sparing TBI using a VMAT technique. Practical radiation oncology Blomain, E. S., Kovalchuk, N. n., Neilsen, E. n., Skinner, L. n., Hoppe, R. T., Hiniker, S. M. 2020

Abstract

Reproductive toxicity is common following total body irradiation and has major quality of life implications for patients. In that context, this is the first report of gonadal-sparing VMAT TBI, successfully delivered in a boy and a girl with aplastic anemia. Both patients' VMAT TBI plans demonstrated improved gonadal sparing versus simulated conventional 2D approach (mean testes dose 0.45 Gy VMAT versus 0.72 Gy 2D; mean ovary dose 0.64 Gy VMAT versus 1.47 Gy 2D). PTV coverage was also improved for both cases with the VMAT plan versus conventional 2D plan (2 Gy D90% versus 1.9 Gy D90%, respectively). Given these dosimetric advantages, the present study can serve as a proof-of-concept for further prospective studies evaluating this technique for wider applications in populations receiving TBI.

View details for DOI 10.1016/j.prro.2020.07.006