A robotically assisted 3D printed quality assurance lung phantom for Calypso. Physics in medicine and biology Capaldi, D. P., Skinner, L. B., Dubrowski, P. n., Zhang, H. n., Xing, L. n., Chuang, C. F., Loo, B. W., Bush, K. K., Fahimian, B. P., Yu, A. S. 2021

Abstract

Purpose:Radiation dose delivered to targets located near the upper-abdomen or in the thorax are significantly affected by respiratory-motion. Relatively large-margins are commonly added to compensate for this motion, limiting radiation-dose-escalation. Internal-surrogates of target motion, such as a radiofrequency (RF) tracking system, i.e. Calypso® System, are used to overcome this challenge and improve normal-tissue sparing. RF tracking systems consist of implanting transponders in the vicinity of the tumor to be tracked using radiofrequency-waves. Unfortunately, although the manufacture provides a universal quality-assurance (QA) phantom, QA-phantoms specifically for lung-applications are limited, warranting the development of alternative solutions to fulfil the tests mandated by AAPM's TG142. Accordingly, our objective was to design and develop a motion-phantom to evaluate Calypso for lung-applications that allows the Calypso® Beacons to move in different directions to better simulate true lung-motion.Methods and Materials:A Calypso lung QA-phantom was designed, and 3D-printed. The design consists of three independent arms where the transponders were attached. A pinpoint-chamber with a buildup-cap was also incorporated. A 4-axis robotic arm was programmed to drive the motion-phantom to mimic breathing. After acquiring a four-dimensional-computed-tomography (4DCT) scan of the motion-phantom, treatment-plans were generated and delivered on a Varian TrueBeam® with Calypso capabilities. Stationary and gated-treatment plans were generated and delivered to determine the dosimetric difference between gated and non-gated treatments. Portal cine-images were acquired to determine the temporal-accuracy of delivery by calculating the difference between the observed versus expected transponders locations with the known speed of the transponders' motion.Results:Dosimetric accuracy is better than TG142 tolerance of 2%. Temporal accuracy is greater than, TG142 tolerance of 100ms for beam-on, but less than 100ms for beam-hold.Conclusions:The robotic QA-phantom designed and developed in this study provides an independent phantom for performing Calypso lung-QA for commissioning and acceptance testing of Calypso for lung treatments.

View details for DOI 10.1088/1361-6560/abebaa

View details for PubMedID 33657537