Longitudinal plaque redistribution during stent expansion AMERICAN JOURNAL OF CARDIOLOGY Maehara, A., Takagi, A., Okura, H., Hassan, A. H., Bonneau, H. N., Honda, Y., Yock, P. G., Fitzgerald, P. J. 2000; 86 (10): 1069-1072

Abstract

The purpose of this study was to clarify the 3-dimensional behavior of plaque during coronary stent expansion. Serial intravascular ultrasound (IVUS) studies, preintervention, and poststenting were evaluated in 32 patients treated with a single-balloon expandable tubular stent. External elastic membrane (EEM), lumen, stent, and plaque + media cross-sectional area were measured at 1-mm intervals through the entire stent as well as proximal and distal reference segments 5 mm from the stent edge. Volumetric calculations were based on Simpson's rule. Overall, the plaque + media volume through the entire lesion did not change during stent expansion (218 +/- 51 vs 217 +/- 47 mm3, p = 0.69). However, EEM and lumen volume increased significantly (EEM volume, 391 +/- 84 vs 448 +/- 87 mm3 [p < 0.0001]; lumen volume, 173 +/- 52 vs 231 +/- 54 mm3 [p < 0.0001]). The change in lumen volume correlated strongly with the change in EEM volume (r = 0.85, p < 0.0001), but poorly with the change in plaque + media volume (r = 0.37, p = 0.03). Plaque + media volume decreased in the midstent zone (59 +/- 14 vs 53 +/- 11 mm3, p = 0.0005), and increased in the distal stent zone (40 +/- 11 vs 44 +/- 9 mm3, p = 0.003), but did not change in either the proximal stent zone or reference segments. The mechanism of stent expansion is a combination of vessel stretch and plaque redistribution, translating disease accumulation from the midstent zone to the distal stent zone.

View details for PubMedID 11074201