Shielding Analysis of a Preclinical Bremsstrahlung X-ray FLASH Radiotherapy System within a Clinical Radiation Therapy Vault. Health physics Rosenstrom, A., Santana-Leitner, M., Rokni, S., Shumail, M., Tantawi, S., Kwofie, J., Dewji, S., Loo, B. W. 2023

Abstract

A preclinical radiotherapy system producing FLASH dose rates with 12 MV bremsstrahlung x rays is being developed at Stanford University and SLAC National Accelerator Laboratory. Because of the high expected workload of 6,800 Gy w-1 at the isocenter, an efficient shielding methodology is needed to protect operators and the public while the preclinical system is operated in a radiation therapy vault designed for 6 MV x rays. In this study, an analysis is performed to assess the shielding of the local treatment head and radiation vault using the Monte Carlo code FLUKA and the empirical methodology given in the National Council on Radiation Protection and Measurements Report 151. Two different treatment head shielding designs were created to compare single-layer and multilayer shielding methodologies using high-Z and low-Z materials. The multilayered shielding methodology produced designs with a 17% reduction in neutron fluence leaking from the treatment head compared to the single layered design of the same size, resulting in a decreased effective dose to operators and the public. The conservative assumptions used in the empirical methods can lead to over-shielding when treatment heads use polyethylene or multilayered shielding. High-Z/Low-Z multilayered shielding optimized via Monte Carlo is shown to be effective in the case of treatment head shielding and provide more effective shielding design for external beam radiotherapy systems that use 12 MV bremsstrahlung photons. Modifications to empirical methods used in the assessment of MV radiotherapy systems may be warranted to capture the effects of polyethylene in treatment head shielding.

View details for DOI 10.1097/HP.0000000000001718

View details for PubMedID 37459481