In Vivo Sodium Imaging of Human Patellar Cartilage With a 3D Cones Sequence at 3 T and 7 T JOURNAL OF MAGNETIC RESONANCE IMAGING Staroswiecki, E., Bangerter, N. K., Gurney, P. T., Grafendorfer, T., Gold, G. E., Hargreaves, B. A. 2010; 32 (2): 446-451

Abstract

To compare signal-to-noise ratios (SNRs) and T*(2) maps at 3 T and 7 T using 3D cones from in vivo sodium images of the human knee.Sodium concentration has been shown to correlate with glycosaminoglycan content of cartilage and is a possible biomarker of osteoarthritis. Using a 3D cones trajectory, 17 subjects were scanned at 3 T and 12 at 7 T using custom-made sodium-only and dual-tuned sodium/proton surface coils, at a standard resolution (1.3 x 1.3 x 4.0 mm(3)) and a high resolution (1.0 x 1.0 x 2.0 mm(3)). We measured the SNR of the images and the T*(2) of cartilage at both 3 T and 7 T.The average normalized SNR values of standard-resolution images were 27.1 and 11.3 at 7 T and 3 T. At high resolution, these average SNR values were 16.5 and 7.3. Image quality was sufficient to show spatial variations of sodium content. The average T*(2) of cartilage was measured as 13.2 +/- 1.5 msec at 7 T and 15.5 +/- 1.3 msec at 3 T.We acquired sodium images of patellar cartilage at 3 T and 7 T in under 26 minutes using 3D cones with high resolution and acceptable SNR. The SNR improvement at 7 T over 3 T was within the expected range based on the increase in field strength. The measured T*(2) values were also consistent with previously published values.