Abstract
Subthreshold retinal laser therapy is efficacious for a variety of retinovascular disorders. Currently, it is unknown which laser parameters can ensure no detectable damage to human retina tissue.One informed physician participant with a normal retina was treated with three levels (75%, 50%, and 25%) of subthreshold 577-nm laser (PASCAL; Topcon, Santa Clara, CA) at 20-millisecond (ms) duration and 100 µm spot size. Several high-resolution retinal imaging modalities, including spectral-domain optical coherence tomography (SD-OCT) and scanning laser ophthalmoscope-adaptive optics (SLO-AO), were used to longitudinally image retinal laser lesions during a 9-month period.SLO-AO and SD-OCT imaging of subthreshold laser therapy in human retina showed no cone cell or RPE damage at all time points during a 9-month period using the 25% threshold power 577-nm laser in the human retina.It is likely that subthreshold laser therapy with 577-nm laser at 20-ms duration in the human retina is safe at the 25% of threshold power level. [Ophthalmic Surg Lasers Imaging Retina. 2016;47:268-275.].
View details for DOI 10.3928/23258160-20160229-10
View details for PubMedID 26985801